RadiForce GX540, 5MP Monochrome LCD Monitor, Part Code No. 0FTD1826A5, 0FTD1826B5, 0FTD1826C5, UPC 690592034752. This pr
FDA device recall Z-2087-2015 · posted 2015-07-16 · Terminated
Recall details
- Recalling firm
- Eizo Corporation
- Reason for recall
- EIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular graphics boards with DisplayPort.
- Action taken
- EIZO sent An Urgent Field Safety Notice dated July 7, 2015, to all customers who purchased the RadiForce RX440, RX650, RX850 and GX540. The notice informs customers that the RadiForce monitors could display abornormally in 6-bit (64 gradations) instead of normal 8-bit (256 gradations) or 10-bit (1024 gradations) while connecting to particular graphics boards with DisplayPort. The notice provides the customers with the actions to be taken and contact information (1-800-800-5202, M-F, 8am-4pm, PST, us.feedback@eizo.com, 5710 Warland Drive, Cypress, CA 90630).
- Root cause
- Device Design
- Status
- Terminated
- Product code
- PGY
- Quantity affected
- 19 units
- Distribution
- Nationwide Distribution including OH, FL, IL, TX, OH, NY, AZ, and NJ.
- Date initiated
- 2015-07-07
- Date posted
- 2015-07-16
- Date terminated
- 2016-11-30
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| 5MP MONOCHROME LCD MONITOR, RADIFORCE GX540 (K130336) | Eizo Nanao Corporation | 2013-03-28 |