R3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 54 MM O.D., REF 71338954, STERILE R, QTY (1), Smith & Nephew, I

FDA device recall Z-2056-2011 · posted 2011-04-22 · Terminated

Recall details

Recalling firm
Smith & Nephew Inc
Reason for recall
During the manufacturing process for several batches of R3 Ceramic Liners, the titanium rings were pressed onto the ceramic component with a higher force than allowed by manufacturing specifications. This has the potential to result in lower than expected strength for the liners.
Action taken
All affected Smith & Nephew Distributors were notified via email and the hospitals were notified by letter delivered by Fed-Ex on 3/11/2011. They were instructed to immediately cease distribution, quarantine affected product for return, if product further distributed, identify customers and immediately notify them of recall. Please contact Mr. David Archer, Group Director, at 701.396.2121, ext. 5373, if you have questions.
Root cause
Employee error
Status
Terminated
Product code
MRA
Quantity affected
571 units
Distribution
AK, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SC, TN, TX, WA, and WI
Date initiated
2011-03-11
Date posted
2011-04-22
Date terminated
2012-02-08
Location
Memphis, TN