Atlas FDA

QuikLYTE Dilution Check: used to monitor and adjust for the dilution ratio on Dimension clinical chemistry systems with

FDA device recall Z-1398-2015 · posted 2015-04-07 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
A positive bias beyond 5% resulting in a dilution check failure or an unnecessary dilution check correction if the positive bias is greater than 1% and less than 5%. The dilution check correction would be detected by QC; however, if not detected by QC, the calibration curve would be decreased and cause elevated sodium, potassium and chloride patient results.
Action taken
An Urgent Medical Device Recall letter, dated February 2015, was provided to all Dimension system customers who received the affected lot to notify them that Siemens confirmed customer complaints of failing dilution checks when using QuikLYTE Dilution Check lot 4MD707. The letter also instructed them to immediately discontinue use and discard all inventory of Dimension QuikLYTE Dilution Check lot 4MD707. Siemens offered a no charge replacement with a non-impacted lot.
Root cause
Under Investigation by firm
Status
Terminated
Product code
CEM
Quantity affected
5981
Distribution
Worldwide Distribution.
Date initiated
2015-02-24
Date posted
2015-04-07
Date terminated
2017-02-01
Location
Newark, DE

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
DADE INDIRECT IMT SYSTEM WITH TCO2 ELECTRODE (K970330)Dade Intl., Inc.1997-04-14