Atlas FDA

Quickie/Zippie powered wheelchair. Q700M wheelchair. A battery operated powered wheelchair, which is controlled by Power

FDA device recall Z-1947-2020 · posted 2020-05-11 · Terminated

Recall details

Recalling firm
Sunrise Medical (US) LLC
Reason for recall
Due to programming errors in the Controller, an increase in the set motor parameters can be made by the end user beyond tested safe limits.
Action taken
An URGENT: MEDICAL DEVICE RECALL letter was sent the firm's dealer. The dealer was instructed on February 4, 2020, to replace the controller unit for the device.
Root cause
Software Design Change
Status
Terminated
Product code
ITI
Quantity affected
1
Distribution
Distributed to US Consignee in MT.
Date initiated
2020-02-13
Date posted
2020-05-11
Date terminated
2021-04-28
Location
Fresno, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Quickie, Zippie (K142457)Sunrise Medical (Us), LLC2014-12-18