Atlas FDA

Quickie Rhythm Power Wheelchair with QC (non-expandable) electronics, Model number EIPW10

FDA device recall Z-1531-2008 · posted 2008-08-23 · Terminated

Recall details

Recalling firm
Sunrise Medical Inc
Reason for recall
Product experiences a longer than expected stopping distance, which may cause injury to the user, objects, or the chair itself.
Action taken
Initial phone contact began on January 11, 2008. Medical device correction letters were also sent to affected dealers.
Root cause
Software design
Status
Terminated
Product code
ITI
Quantity affected
141 units total
Distribution
Product was distributed to 126 retailers nationwide throughout the USA, as well as Canada and Australia.
Date initiated
2008-01-11
Date posted
2008-08-23
Date terminated
2009-11-16
Location
Fresno, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
QUICKIE RHYTHM & GROOVE SERIES POWER WHEELCHAIRS W/QUICKIE DELPHI CONTROLS (K060531)Sunrise Medical2006-03-17