Quick Shield with Snappy Safety Holder, Greiner bio-one. The VACUETTE QUICKSHIELD with SNAPPY Tube Holder is intended to

FDA device recall Z-0179-2012 · posted 2011-11-09 · Terminated

Recall details

Recalling firm
Greiner Bio-One North America, Inc.
Reason for recall
Reports of the Vacuette QuickShield Safety Holders breaking during use due to a crack at the holder hub.
Action taken
Greiner Bio-One sent a "QUICKSHIELD WITH SAFETY HOLDER" notification letter dated August 11, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to stop distributing and using the affected products. A Product Disposition Form was attached for customers to complete and return for replacements. Contact the firm at 888-266-3883 for questions concerning this recall.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
FMI
Quantity affected
443,950 pieces
Distribution
Nationwide Distribution-USA (nationwide) including the states of CO, IL, KY, NY, PA and VA.
Date initiated
2011-08-11
Date posted
2011-11-09
Date terminated
2012-03-20
Location
Monroe, NC

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
VACUETTE QUICKSHIELD WITH SNAPPY TUDE HOLDER (K102774)Greiner Bio-One North America, Inc.2010-10-15