FDA device recall Z-1399-2010 · posted 2010-04-15 · Terminated

Recall details

Recalling firm
Orthohelix Surgical Designs Inc
Reason for recall
The adapters are able to come apart when a retaining ring in the proximal end of the adapter can dislodge from the retaining ring groove and the two ends of the adapter fall apart and are unable to be reassembled without special tools and knowledge of the instrument. Also the adapters have a wobble between the two ends. This is caused when the distal and proximal ends of the adapter assembly were
Action taken
Two "IMPORTANT RECALL NOTICE" letters, one dated October 16, 2009, and the second letter dated November 19, 2009, were sent to the customers. The letters described the product, problem and action to be taken by the customers. The customer are requested to discontinue use of the adapters and return them to the firm, Orthohelix Surgical Designs. The firm will replace the adapters as soon as possible. Should you have any questions about the systems please contact Kristin Wolff at 330-247-1445 or Jim Bragg at 330-247-1444.
Root cause
Other
Status
Terminated
Product code
LXH
Quantity affected
127 pieces
Distribution
Nationwide distribution: CA, CO, FL, GA, IL, KY, MD, MI, MO, MS, NC, NJ, NM, NY, OH, OR, PA, SD, TN, TX, VA, WA , and WI.
Date initiated
2009-10-16
Date posted
2010-04-15
Date terminated
2012-04-11