Quantimetrix Urine Chemistry Control Set Dipper Urinalysis Dipstick Testing Levels 1 & 2 6 X 15 mL-IVD intended as a con
FDA device recall Z-0138-2025 · posted 2024-10-22 · Open, Classified
Recall details
- Recalling firm
- Mckesson Medical-Surgical Inc. Corporate Office
- Reason for recall
- The Quantimetrix Dipper Urinalysis Dipstick control was shipped without the validated summer packaging and may have been exposed to temperatures above 25C. This may lead to delay in patients' results.
- Action taken
- McKesson issued Urgent Product Recall letter via FedEx Overnight service scheduled for delivery on 9/17/24 to one consignee. Letter states reason for recall, health risk and action to take: 1. Immediately examine your inventory and quarantine all product subject to this urgent product recall detailed in the affected products list (Attachment #1). 2. Please complete, sign, and return the enclosed response form (Attachment #2 - Customer Response Form) as soon as possible, even if you do not have any product on hand. Return completed signed form MMSQARecalls@McKesson.com. 3. If you have any product, destroy using your local procedures and document the destruction on attached customer response form (Attachment #2 - Customer Response Form). Return completed signed form MMSQARecalls@McKesson.com. 4. If you have further distributed this product, please notify them immediately of this urgent product recall. You are encouraged to include a copy of this notification letter in your communication. If you have questions regarding this notification, please contact McKesson MedicalSurgical directly by phone at (800) 688-8840. Representatives are available Monday -Friday, 8:00 am - 5:00 pm.
- Root cause
- Packaging process control
- Status
- Open, Classified
- Product code
- JJW
- Quantity affected
- 1 BOX
- Distribution
- US Nationwide distribution in the state of DC.
- Date initiated
- 2024-09-17
- Date posted
- 2024-10-22
- Location
- Richmond, VA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| QUANTIMETRIX URINE DIPSTICK CONTROL (K874890) | Quantimetrix Medical Industries | 1988-01-21 |