Atlas FDA

Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas

FDA device recall Z-0608-2013 · posted 2012-12-21 · Terminated

Recall details

Recalling firm
Radiometer America Inc
Reason for recall
The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.
Action taken
Radiometer sent Field Action Notes regarding the affected product. The note identified the product, the problem and the action to be taken by the customer. Instead of scanning the barcorde on the insert to enter the QC data into the anyalyzer, customers were instructed to manually key in the QUALICHECK type, lot number, and correct control range. For questions regarding this recall call 440-871-8900.
Root cause
Process control
Status
Terminated
Product code
JJE
Quantity affected
840 Ampules
Distribution
Distributed to KY, NC, and NY.
Date initiated
2006-06-06
Date posted
2012-12-21
Date terminated
2013-02-05
Location
Westlake, OH