QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 300
FDA device recall Z-2262-2013 · posted 2013-09-20 · Terminated
Recall details
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Reason for recall
- The manufacturer has conducted a review of the results of product testing performed during the last twelve months and has identified specific lots of product which may not meet specifications for bond strength on the inlet and/or outlet connector.
- Action taken
- Maquet sent a Urgent Product Recall letter/Response Form, dated 8/15/2013 via Fed Ex. On 8/26/2013 Maquet sent a letter to their customers informing them that all shipments of oxygenators and stainless steel hold clamps have been placed on a temporary hold due to an agency review in connection with FDA clearance. Customer service may be contacted at 1-888-627-8383, (option 2 followed by option 2) Monday though Friday between the hours of 8:00 am and 6:00 pm EST. Maquet issued a Certificate of Medical Neccesity (CMN) on September 17, 2013 to notify Quadrox customers that they may continue to receive and use their current inventory of Quadrox oxygenators and clamps by acknowledging the conditions provided in the CMN.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- DTZ
- Quantity affected
- 1,182 units
- Distribution
- Worldwide Distribution, including Nationwide (US) and foreign countries.
- Date initiated
- 2013-08-15
- Date posted
- 2013-09-20
- Date terminated
- 2017-07-11
- Location
- Wayne, NJ
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| QUADROX-I MICROPORPOROUS MEMBRANE OXGENATOR SERIES AND QUDROX-ID DIFFUSION MEMBRANE OXYGENATORS SERIES (K112360) | Maquet Cardiopulmonary, AG | 2011-09-15 |
| QUADROX-I PEDIATRIC MICROPOROUS MEMBRANE OXYGENATOR WITH INTEGRATED ARTERIAL FILTER WITH SOFTLINE / BIOLINE COATING (K103191) | Maquet Cardiopulmonary, AG | 2011-03-08 |
| QUADROX-I SMALL ADULT MICROPOROUS MEMBRANE OXYGENATOR; W/ BIOLINE COATING AND W/ INTEGRATED ARTERIAL FILTER W/ BIOLINE C (K093522) | Maquet Cardiopulmonary, AG | 2009-12-07 |
| QUADROX-I SMALL ADULT WITH AND WITHOUT INTERGRATED ARTERIAL FILTER WITH SOFTLINE COATING, MODELS HMO 50000, HMO 51000 (K090689) | Maquet Cardiopulmonary, AG | 2009-10-29 |
| QUADROX-I ADULT MICROPOROUS MEMBRANE OXYGENATOR WITH BIOLINE COATING, MODEL BEHMO70000,BEQ-HMO70000 (K090511) | Maquet Cardiopulmonary, AG | 2009-06-12 |
| QUADROX-I ADULT WITH AND WITHOUT INTEGRATED ARTERIAL FILTER WITH SOFTLINE COATING, MODEL# HMO 70000, HMO 71000 (K082117) | Maquet Cardiopulmonary, AG | 2009-02-18 |