QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 300
FDA device recall Z-0160-2013 · posted 2012-10-26 · Terminated
Recall details
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Reason for recall
- It has come to firm's attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units shipped.
- Action taken
- Maquet sent a Field Safety Notices and Confirmation Forms dated June 13, 2012 to all affected customers. A Follow-up communication dated July 20, 2012 was mailed to all affected US customers. The Field Safety Notice identifies the problem, affected product, risk factors and actions to be taken by the customer. Customers were instructed to complete the Confirmation form and return in the self-addressed prepaid envelop. For questions call 201-995-8968. Maquet issued an updated Field Safety Notice/Field Safety Notice Confirmation Form dated 2/14/13 via Fed Ex on 2/18/2013. Maquet sent out an Updated Field Safety Notice/field Safety Notice Confirmation Form dated 4/4/2013 in order to notify customers they would be providing safety clamps.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- DTZ
- Quantity affected
- 23,203 units
- Distribution
- US Nationwide Distribution
- Date initiated
- 2012-06-12
- Date posted
- 2012-10-26
- Date terminated
- 2017-09-29
- Location
- Wayne, NJ
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| QUADROX-I MICROPORPOROUS MEMBRANE OXGENATOR SERIES AND QUDROX-ID DIFFUSION MEMBRANE OXYGENATORS SERIES (K112360) | Maquet Cardiopulmonary, AG | 2011-09-15 |
| QUADROX-I PEDIATRIC MICROPOROUS MEMBRANE OXYGENATOR WITH INTEGRATED ARTERIAL FILTER WITH SOFTLINE / BIOLINE COATING (K103191) | Maquet Cardiopulmonary, AG | 2011-03-08 |
| QUADROX-I SMALL ADULT MICROPOROUS MEMBRANE OXYGENATOR; W/ BIOLINE COATING AND W/ INTEGRATED ARTERIAL FILTER W/ BIOLINE C (K093522) | Maquet Cardiopulmonary, AG | 2009-12-07 |
| QUADROX-I SMALL ADULT WITH AND WITHOUT INTERGRATED ARTERIAL FILTER WITH SOFTLINE COATING, MODELS HMO 50000, HMO 51000 (K090689) | Maquet Cardiopulmonary, AG | 2009-10-29 |
| QUADROX-I ADULT MICROPOROUS MEMBRANE OXYGENATOR WITH BIOLINE COATING, MODEL BEHMO70000,BEQ-HMO70000 (K090511) | Maquet Cardiopulmonary, AG | 2009-06-12 |
| QUADROX-I ADULT WITH AND WITHOUT INTEGRATED ARTERIAL FILTER WITH SOFTLINE COATING, MODEL# HMO 70000, HMO 71000 (K082117) | Maquet Cardiopulmonary, AG | 2009-02-18 |