Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the s

FDA device recall Z-0267-2013 · posted 2012-11-08 · Terminated

Recall details

Recalling firm
Medacta Usa Inc
Reason for recall
Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.
Action taken
Medacta USA, sent a Urgent Safety Notification letter dated October 13, 2008, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. The letter informed the customers of the lot numbers on recall and the reference numbers. The recall was initiated because the offset broach handles contain a lever mechanism that is connected to the locking axis by means of a metallic wire. The rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches. For further questions please call (858) 705-0350.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JDI
Quantity affected
12 units
Distribution
Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.
Date initiated
2008-10-13
Date posted
2012-11-08
Date terminated
2012-11-09
Location
Camarillo, CA