FDA device recall Z-2246-2023 · posted 2023-07-27 · Open, Classified

Recall details

Recalling firm
Qiagen GmbH
Reason for recall
Identified a decreased performance reliability rate, Run abortions could cause delayed diagnosis since sample testing would have to be repeated
Action taken
Qiagen issued a customer letter (Urgent Medical Device Correction (REF 691223) or Urgent Field Safety notice (REF 691214) via e-mail on 5/19/23. Letter states reason for recall, health risk and action to take: Do not use the remaining stock of cartridges of LOTs 175010704 and 175011354. Dispose of it immediately in accordance to your national and local safety and environmental regulations. " Please contact QIAGEN Technical Services for a free-of-charge replacement. " Review this notice with your laboratory/medical director. " Forward this information to all individuals and departments within your organization using the above listed kits. If you are not the end user, please forward this notice to the product end user. " Complete the Acknowledgement of Receipt Form attached to this letter by 02 June 2023. you have any questions or concerns, please contact your local QIAGEN Technical Services Department through any of the following: Telephone: 800 362 7737 Email: TechService-NA@qiagen.com
Root cause
Process control
Status
Open, Classified
Product code
QLT
Quantity affected
4155 kits
Distribution
DC, AL, CA, FL, GA, ME, NC, NY, OR, TX Foreign: DE, AE, AT, BE, BG, CH, CL, CY, CZ, DE, DK, ES, FI, FR, GB, GR, GT, HK, HU, ID, IN, IT, JO, MC, MU, MY, NL, NA, PL, PT, RO, SE, SK, TH, TN, TW, UY, YT,
Date initiated
2023-05-19
Date posted
2023-07-27
Location
Hilden, Germany