Puritan Bennett 840 ventilator systems software part number 4-070212-85, revision AB-AG. The intended use of the product

FDA device recall Z-0611-2014 · posted 2014-01-03 · Terminated

Recall details

Recalling firm
Covidien
Reason for recall
Covidien is conducting a voluntary field corrective action on Puritan Bennett 840 ventilator systems with certain software revisions in response to customer reports of ventilator diagnostic code XB0069 in which the device stops mechanical ventilation due to a software error.
Action taken
Covidien sent an Urgent Medical Device Voluntary Field Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to continue using their Puritan Bennett 840 ventilators until it is feasible to install the software update discussed in the letter. Covidien has developed a software solution to address this issue. Upon availability, there will be several ways to install the software update. To initiate the process for updating your software and to select the method that best suits your facility, please visit our software update management portal at www.PB840technicalupdate.com and follow the instructions to register your ventilators. If you need further assistance contact our Technical Support Department @ 1-800-255-6774, option 4, then option 1. For questions regarding this recall call 800-255-6774.
Root cause
Software design
Status
Terminated
Product code
CBK
Quantity affected
16,923 US, 39,372 OUS
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Austria, Estonia, Belgium, Spain, Italy, Poland, Bulgaria, Finland, Portugal, Switzerland, France, Lithuania, Romania, Cyprus, Great Br
Date initiated
2013-12-16
Date posted
2014-01-03
Date terminated
2016-11-29
Location
Boulder, CO

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
MODEL 840 VENTILATOR WITH PROPORTIONAL ASSIST VENTILATION PLUS SOFTWARE OPTION (K053388)Puritan Bennett Corp.2006-04-04
800 SERIES EASYNEB NEBULIZER (K020357)Nellcor Puritan Bennett, Inc.2002-08-07
PURITAN-BENNETT 840 VENTILATOR SYSTEM WITH NEOMODE OPTION (K001646)Puritan Bennett Corp.2000-10-20
PURITAN-BENNETT VENTILATOR SYSTEM WITH ET/TRACHEOSTOMY TUBE COMPENSATION, TC, OPTION, MODEL 840 (K993071)Puritan Bennett Corp.1999-11-24
PURITAN-BENNETT 840 VENTILATORY SYSTEM WITH BILEVEL OPTION (K984535)Puritan Bennett Corp.1998-12-28
NELLCOR PURITAN BENNETT, 840 VENTILATOR SYSTEM (MODEL 840) (K970460)Puritan Bennett Corp.1998-04-30