Puritan-Bennett 800 Series Ventilator Backup Power Source (BPS) Model 802, Puritan-Bennett Corporation, A subsidiary of

FDA device recall Z-1692-2008 · posted 2008-09-16 · Terminated

Recall details

Recalling firm
Covidien Limited
Reason for recall
Wiring in battery back-up power supply may short and cause thermal damage to ventilator.
Action taken
All customers were notified by Federal Express letter and were contacted by phone through field service engineers. They were told to examine their inventory for any installed or spare part battery assemblies and to remove and quarantine these parts. Any ventilators with installed affected battery assemblies were to be taken out of service.
Root cause
Process control
Status
Terminated
Product code
CBK
Quantity affected
6 Ventilators and 151 factory replacement units
Distribution
Nationwide, including one VA facility in IL and two military facilities in NC and NY. No foreign distribution.
Date initiated
2008-03-21
Date posted
2008-09-16
Date terminated
2008-10-29
Location
Boulder, CO