Atlas FDA

PTS Panels CHOL+HDL+GLU Panel test strips for professional use with CardioChek P-A analyzers, Manufactured by Polymer Te

FDA device recall Z-1012-2010 · posted 2010-03-05 · Terminated

Recall details

Recalling firm
Polymer Technology Systems, Inc.
Reason for recall
Possible under recovery of cholesterol, HDL and glucose. A loss of recoverable cholesterol and HDL cholesterol is a manageable to moderate health risk. The clinical applications analysis shows that lot 1891 shows such a loss of activity. A loss of recoverable Glucose measurement is a moderate to high risk, based upon the degree of under-recovery, and the frequency of occurrence.
Action taken
Consignees were notified by letter dated 7/28/09 to discontinue use of, and to destroy, the reagent and to contact the firm for replacement product. For questions, please contact the company Customer Service toll-free at 877-870-5610 or + 1-317-870-5610 (outside the US).
Root cause
Other
Status
Terminated
Product code
CGA
Quantity affected
962
Distribution
Nationwide.
Date initiated
2009-08-04
Date posted
2010-03-05
Date terminated
2010-10-20
Location
Indianapolis, IN