Psi-Tec Tubing, 12'; 10/box; Part Number PT-5558; distributed by Mentor Texas, LP, Irving, TX 75038.

FDA device recall Z-1790-2008 · posted 2008-09-05 · Terminated

Recall details

Recalling firm
Mentor Texas, Inc
Reason for recall
During packaging integrity validation, it was found that various product packages could fail, compromising the sterility of the devices.
Action taken
A Customer Letter was issued to all consignees by the firm on 4/21/2008. Consignees were informed that a recent testing of these products discovered a potential defect in the packaging configuration, which could result in a loss of package integrity. Consignees were asked to review inventory and dispose of product and contact firm for credit.
Root cause
Other
Status
Terminated
Product code
FPK
Quantity affected
18,684 boxes.
Distribution
Class II Recall - Worldwide Distribution ---- including USA, Puerto Rico and countries of Argentina, Austria, Bermuda, Brazil, Canada, China, Costa Rica, Cyprus, Denmark, England, Germany, Greece, Gua
Date initiated
2008-04-21
Date posted
2008-09-05
Date terminated
2010-07-29
Location
Irving, TX