ProVue Instrument Driver for JResultNet (ProVue.jrm) v. 1.5.3 and below JResultNet is a stand-alone software package tha
FDA device recall Z-2637-2014 · posted 2014-09-12 · Terminated
Recall details
- Recalling firm
- Data Innovations, LLC
- Reason for recall
- Potential issues with the ProVue driver have been identified which may cause data to be misassociated with patient results under a given set of circumstances.
- Action taken
- Data Innovations LLC sent an Important Medical Device Information letter on August 15, 2014, to consignees for the JResultNet using ProVue Instrument Driver JResultNet (provue.jrm) v. 1.5.3 and below. Potential issues with the ProVue driver have been identified which may cause data to be mis-associated with patient results under a given set of circumstances. The user on the ProVue transmits the same results twice back to back by clicking the Send to Host button twice. The ProVue sends two ASTM messages back to back before ending transmission (sending an EOT character) The solution was to disconnect JResultNet from the ProVue instrument and apply the patch to driver provue.jrm when it is available. If you have further questions regarding this notice, please feel free to contact us at northamerica-support@datainnovations.com or 239-931-6004 Toll-Free (from the US and Canada): 1-800-332-0499.
- Root cause
- Software design
- Status
- Terminated
- Product code
- JQP
- Quantity affected
- 84
- Distribution
- Nationwide Distribution including AL, AZ, CA, CT, DC, FL, GA, IN, LA, MI, MO, NC, NH, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, and WA.
- Date initiated
- 2014-08-15
- Date posted
- 2014-09-12
- Date terminated
- 2015-03-04
- Location
- Fort Myers, FL