Progressive Medical PMI Endoscopic Dissector with Kittner Tip (5mm x 40cm), Reorder Number: PMILK40, Product Sterile in
FDA device recall Z-0159-2013 · posted 2012-10-26 · Terminated
Recall details
- Recalling firm
- Progressive Medical Inc
- Reason for recall
- The tip of the product could come loose or unravel.
- Action taken
- The recalling firm sent a letter dated October 28, 2011 to the customer who reported the problem and calls were initiated 6/26/12 to the remainder of the customers. The reason for recall was explained and the customer were asked if any of the lot remained in their inventory. If inventory remained, the recalling firm sent them a return goods authorization number and informed them the product would be replaced. For questions call 1-800-969-6331.
- Root cause
- Packaging
- Status
- Terminated
- Product code
- GDI
- Quantity affected
- 124 boxes
- Distribution
- US Nationwide Distribution in the states of: CO, IL, IN, KS, LA, MA, MN, MO, NE, OH, OK, TN, TX, VA, and WI.
- Date initiated
- 2011-10-28
- Date posted
- 2012-10-26
- Date terminated
- 2012-10-26
- Location
- Saint Louis, MO