Product REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery Unit
FDA device recall Z-0710-2024 · posted 2024-01-12 · Open, Classified
Recall details
- Recalling firm
- Maquet Cardiovascular, LLC
- Reason for recall
- Express chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
- Action taken
- On December 13, 2023, the firm notified customers through "URGENT MEDICAL DEVICE CORRECTION" letters. Customers were instructed to examine their inventory to determine if they had any affected lots of the Atrium Express Dry Suction Dry Seal Chest Drains. The correction only affects the sterile water syringe; do not use this syringe. If customers do not need visualization of active pneumothorax, they may keep the affected Express chest drain and set up without water. If customers wish to use the Express chest drain as intended, they may replace the affected sterile water syringe with a new syringe filled with sterile water using aseptic technique. Customers also have the option of returning affected Express chest drains to the firm for credit. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m.(Eastern Time Zone).
- Root cause
- Material/Component Contamination
- Status
- Open, Classified
- Product code
- CAC
- Quantity affected
- (7,501 cases/45,006 devices US); no OUS
- Distribution
- Worldwide distribution: US (nationwide) and OUS (Foreign) to countries of: Australia, Brazil, Canada, Luxembourg, Netherlands, Norway, Spain and UK
- Date initiated
- 2023-12-13
- Date posted
- 2024-01-12
- Location
- Wayne, NJ
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|