Atlas FDA

Product REF 4000-100N, DRAIN, EXPRESS SINGLE W/AC, Single Collection

FDA device recall Z-0709-2024 · posted 2024-01-12 · Open, Classified

Recall details

Recalling firm
Maquet Cardiovascular, LLC
Reason for recall
Express chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
Action taken
On December 13, 2023, the firm notified customers through "URGENT MEDICAL DEVICE CORRECTION" letters. Customers were instructed to examine their inventory to determine if they had any affected lots of the Atrium Express Dry Suction Dry Seal Chest Drains. The correction only affects the sterile water syringe; do not use this syringe. If customers do not need visualization of active pneumothorax, they may keep the affected Express chest drain and set up without water. If customers wish to use the Express chest drain as intended, they may replace the affected sterile water syringe with a new syringe filled with sterile water using aseptic technique. Customers also have the option of returning affected Express chest drains to the firm for credit. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m.(Eastern Time Zone).
Root cause
Material/Component Contamination
Status
Open, Classified
Product code
CAC
Quantity affected
(34,356 cases/206,136 devices US); (3,902 cases/23,412 devices OUS)
Distribution
Worldwide distribution: US (nationwide) and OUS (Foreign) to countries of: Australia, Brazil, Canada, Luxembourg, Netherlands, Norway, Spain and UK
Date initiated
2023-12-13
Date posted
2024-01-12
Location
Wayne, NJ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
EXPRESS AND OASIS CHEST DRAIN (K043140)Atrium Medical Corp.2004-12-08