Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculati
FDA device recall Z-2509-2026 · posted 2026-06-18 · Open, Classified
Recall details
- Recalling firm
- Staar Surgical AG
- Reason for recall
- Due to a software coding issues for calculating the implant orientation diagram printout for the implantation for toric implantable collamer lenses.
- Action taken
- On 04/20/2026, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL" Letter informing customers that STAAR has confirmed that during that period, the printing of the IOD within the Stella Patient View, Order Confirmation, or Shopping Cart, when performed locally (i.e. by doctor/clinic), was under certain circumstances producing an incorrect diagram. The incorrect diagram was produced in certain circumstances when users created multiple Toric lens line items with different axis values on the same patient eye, such as with backup or secondary lens reservations. The root cause has been identified as a software coding error in Stella 2.0 that did not assign a unique identifier to each reservation under a patient s eye. STAAR has modified the software code to assign a unique identifier, thereby preventing this issue when multiple reservations exist. Any IOD printed from Stella 2.0 after 5:30pm PDT on April 16, 2026 would not have been affected by this issue. For Customer's awareness: We are informing you so that a proper clinical decision can be made regarding patient management and follow-up, based on your assessment of the clinical outcome. Please immediately complete and return the Recall Response Form acknowledging receipt of this recall and report any adverse events associated with this event. If a Toric EVO/EVO+ device were implanted using an incorrect IOD printout, there is the potential for the following adverse events: " Refractive surprise " Blurred vision " Glares/Halos " Other visual disturbances Repositioning or removal of the lens may need to be considered in these cases. For Questions - contact STAAR Sales Representative or Customer Service at customerservice@staar.com or by phone at 1-800-352-7842 Monday through Friday, 6:00am to 5:00pm PDT.
- Root cause
- Software design
- Status
- Open, Classified
- Product code
- MTA
- Quantity affected
- 543
- Distribution
- Worldwide - US Nationwide distribution including in the states and territory of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM,
- Date initiated
- 2026-04-20
- Date posted
- 2026-06-18
- Location
- Nidau, Switzerland
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.