Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10
FDA device recall Z-1298-2026 · posted 2026-02-05 · Open, Classified
Recall details
- Recalling firm
- Wilson-Cook Medical Inc.
- Reason for recall
- Nasal feeding tube packaged without the nasal transfer tube component.
- Action taken
- On January 5, 2026, Cook Endoscopy issued a Urgent Medical Device Recall Notification to affected consignees via E-Mail. Cook Endoscopy asked consignees to take the following actions: 1. Examine inventory immediately to determine if you have affected product(s). If so, quarantine any affected products that remain unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a signed copy of the Acknowledgement and Receipt Form. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand and email the completed form to ausrecalls@cookmedical.com. 4. Please share this notice with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected products have been transferred.
- Root cause
- Employee error
- Status
- Open, Classified
- Product code
- KNT
- Quantity affected
- 60 units
- Distribution
- International distribution to the country of Australia.
- Date initiated
- 2025-12-15
- Date posted
- 2026-02-05
- Location
- Winston Salem, NC
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|