prodisc C SK, Total cervical disc replacement, medical device, Model/Catalog Number: PDSM5

FDA device recall Z-3212-2024 · posted 2024-09-24 · Open, Classified

Recall details

Recalling firm
Centinel Spine, Inc.
Reason for recall
During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration is considered a foreign contaminate that may not have been fully evaluated in the manufacturing validations.
Action taken
Customers were contacted via telephone beginning 7/18/24. The Centinel Quality team has identified a potential nonconformance on three specific lots in the prodisc c product family. We are requesting the return of the inventory that is assigned to you. We will provide replacement inventory proactively to support the business. Thank you for your support. In addition, an URGENT - MEDICAL DEVICE Recall Instructions for PRODISC C IMPLANTS notification letter was sent to the distributor. Required Action: Please retrieve the identified devices and return the affected devices to Centinel Spine Depth of Recall: This recall should be carried out to the hospital level, including Distributors and Consigned locations Please acknowledge that you have read the email and enable the Return Receipt. This is necessary to ensure traceability as part of the recall process. If you have any questions, please contact Jessica Staub 484.887.8803.
Root cause
Process control
Status
Open, Classified
Product code
MJO
Quantity affected
12 units
Distribution
US Nationwide distribution.
Date initiated
2024-07-18
Date posted
2024-09-24
Location
West Chester, PA