Premier brand Toxins A & B, diagnostic test kits, one kit per pouch, one pouch per shipping box.

FDA device recall Z-0464-03 · posted 2003-01-23 · Terminated

Recall details

Recalling firm
Meridian Bioscience Inc
Reason for recall
Due to defective packaging, the firm cannot assure product stability through the labeled expiration date.
Action taken
The firm mailed notices to their affected customers on 12/26/2002. The consignees were instructed to visually and physically inspect the seals on each microwell pouch for gaps. If properly sealed, the use of the microwells may be continued. Test wells in pouches with defective seals are to be quarantined and destroyed. The customers are asked to report back to the firm. All test microwell plates in defective pouches will be replaced by the firm.
Root cause
Other
Status
Terminated
Product code
MCB
Quantity affected
1,409 kits
Distribution
The kits were distributed to end-users and distributors located nationwide; and to foreign distributors located in China, Guatemala, New Zealand, Hong Kong, Mexico, Great Britain, Belgium, Italy, Braz
Date initiated
2002-12-26
Date posted
2003-01-23
Date terminated
2004-03-29
Location
Cincinnati, OH