Prelude Short Sheath Introducer, 7F-SMT, 4 cm, REF No: PSS-7F-4-038MT, Sterile EO, Merit Medical, South Jordan, Utah. In

FDA device recall Z-0941-2011 · posted 2011-01-28 · Terminated

Recall details

Recalling firm
Merit Medical Systems, Inc.
Reason for recall
Introducer tips may detach and embolize during use, or the tip material may elicit a thrombogenic response.
Action taken
Merit Medical Systems, Inc., issued an "Urgent Product Recall Notice" dated December 17, 2010, and contacted all customers (sales reps and authorized distributors) via email, phone and/or fax. The notice described the product, problem and action to be taken by the customers (Sales reps/Distributors). The customers were instructed to contact all their customers to identify, quarantine and return all unused inventory and extend the recall to any of their customers who may have received the affected products. The customers were instructed to complete and return the Manufacturer RECALL NOTIFICATION and Product Retrieval Form via fax to 1-801-208-3378 and send original to: the Manager, Customer Service, Merit Medical Systems, Inc., 1600 West Merit Parkway, South Jordan, UT 84095-2416. Questions should be directed to the Senior Product Manager at 801-208-4344 or the Manager, Customer Service at 801-208-4365.
Root cause
Component change control
Status
Terminated
Product code
DYB
Quantity affected
378 units
Distribution
Nationwide Distribution
Date initiated
2010-12-17
Date posted
2011-01-28
Date terminated
2011-07-11
Location
South Jordan, UT