Atlas FDA

PrecisePLAN 2.10 Treatment Planning System

FDA device recall Z-1406-2010 · posted 2010-04-16 · Terminated

Recall details

Recalling firm
Elekta, Inc.
Reason for recall
Error in dose rates which were calculated at exactly twice the expected values.
Action taken
As described in a prior related submission, Phase I - "Important Notice A285, Scrolling Selected Frame Dose Calculator Error" dated March 15, 2006 was distributed to affected customers to inform them of the potential and what steps should be taken. Phase 2 -- Software release PrecisePLAN 2.11 was released July 11, 2006 to correct this problem. All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. Phase 3 -- All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. The company is directing questions to the local Electa representative or to their offices in the U.K. at telephone 44 (0)-1293-654200.
Root cause
Software design
Status
Terminated
Product code
MUJ
Quantity affected
40 units
Distribution
Worldwide: IL, NY, PA, TN, Canada, China, Czech Republic, Germany, India, Israel, Italy, Mexico, Poland, Russia, Spain, Switzerland
Date initiated
2006-03-15
Date posted
2010-04-16
Date terminated
2010-12-30
Location
Norcross, GA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
PRECISEPLAN 2.0 (K022411)Precision Therapy Intl., Inc.2002-09-20