FDA device recall Z-1432-2010 · posted 2010-04-23 · Terminated

Recall details

Recalling firm
Elekta, Inc.
Reason for recall
In some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table.
Action taken
Mandatory Field Change Order (FCO) 528 053 "Check for zinc plated Precise Table 'Z' mechanism fixation bolts", dated 1/11/2007 was distributed to all affected customers. The Change Order identified the product, stated the reason for the FCO, and the procedure to check and resolve the problem identified.
Root cause
Other
Status
Terminated
Product code
JAI
Quantity affected
185 units
Distribution
Worldwide Distribution -- USA, Puerto Rico, Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Ira
Date initiated
2007-01-11
Date posted
2010-04-23
Date terminated
2010-12-28
Location
Norcross, GA