Atlas FDA

PowerPICC SOLO Catheter with Tip Location System (TLS) Stylet 4F Nurse Full Tray with Microintroducer (with Lidocain) Th

FDA device recall Z-2049-2015 · posted 2015-07-08 · Terminated

Recall details

Recalling firm
Bard Access Systems
Reason for recall
Bard Access Systems is recalling PowerPICC Solo Catheter Kits due to incorrect expiration dates on the primary labeling.
Action taken
BARD sent an Urgent Notification letter dated May 27, 2015, to all affected consignees via FedEx. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed that they may choose to return the affected product to Bard Access Systems or use with the understanding that the correct expiration date is November 2015 (2015-11). Customers with questions were instructed to call 1-800-290-1689. For questions regarding this recall call 801-522-5974.
Root cause
Labeling Change Control
Status
Terminated
Product code
LJS
Distribution
Nationwide Distribution
Date initiated
2015-05-27
Date posted
2015-07-08
Date terminated
2016-02-10
Location
Salt Lake City, UT