Powerheart G3 Automated External Defibrillators (Non-Wearable) and include the following models: Powerheart 9300A, 9300E

FDA device recall Z-1057-2012 · posted 2012-02-27 · Terminated

Recall details

Recalling firm
Cardiac Science Corporation
Reason for recall
The affected Automated External Defibrillators (AEDs) contains a circuit board manufactured with a component that may fail unexpectedly due to a supplier manufacturing defect. If the component were to fail during a rescue attempt, the AED may not deliver defibrillation therapy.
Action taken
Cardiac Science sent an "URGENT-Voluntary Medical Device Recall" letter dated January 13, 2012 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. Customers are advised to immediately return the device to the manufacturing location for repair. The affected devices will receive a hardware correction, and the same serial number device returned to the consignee in most cases. Customers in the US can call 1-888-402-2484 and customers outside the US can call 1-425-402-2482 or email at aed210@cardiacscience.com to arrange delivery of the shipping materials. Customers can also visit www.cardiacscience.com/aed210 for more information about AEDs affected by this recall.
Root cause
Process control
Status
Terminated
Product code
MKJ
Quantity affected
11,019 units (6142 units in the US and 4877 units international)
Distribution
Worldwide distribution: USA (nationwide) and Puerto Rico; and to the following countries: Australia, Belgium, Brazil, Canada, China, Colombia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France
Date initiated
2012-01-13
Date posted
2012-02-27
Date terminated
2013-09-27
Location
Bothell, WA