Atlas FDA

Power cord for Optional Thermal Printer (For use with Model 3300 COC), Oximetrix 3 SO2/CO Computer, Q-Vue CCO Computer a

FDA device recall Z-0143-2010 · posted 2009-11-09 · Terminated

Recall details

Recalling firm
Hospira Inc
Reason for recall
Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock hazard.
Action taken
Hospira initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and a letter to consignees was sent via UPS. The firm will replace affected cords and effectiveness checks will be conducted through tracking of servicing.
Root cause
Component design/selection
Status
Terminated
Product code
dxg
Quantity affected
204,999 power cords distributed for all devices
Distribution
Worldwide distribution.
Date initiated
2009-08-11
Date posted
2009-11-09
Date terminated
2012-04-04

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
MODEL 3000 OPTICATH COMPUTER (K853018)Oximetrix, Inc.1986-01-13