Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile

FDA device recall Z-0918-2025 · posted 2025-01-15 · Open, Classified

Recall details

Recalling firm
GE Medical Systems, LLC
Reason for recall
There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.
Action taken
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 12/6/24 was sent to customers. Actions to be taken by Customer/User: You can continue to use your Portrait" Mobile Monitoring Solution v1.1 product, since all potentially impacted systems were recently installed and the safety issue does not manifest until after 425 days of continuous use. Please follow the annual Maintenance Schedule described in Service Manual (PN 5929144) for your Portrait" Mobile to reboot the system. Rebooting the system by following this procedure resets the continuous days of use to zero. Please complete and return the attached Acknowledgement Form to FMI.36165@gehealthcare.com. After the Portrait Core Services v1.1 software is updated and corrected by GE HealthCare, please discontinue usage of any Portrait Core Services v1.1 software versions 1.1.0 or 1.1.1 and destroy any software media containing those versions. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Root cause
Software design
Status
Open, Classified
Product code
MWI
Quantity affected
7 systems
Distribution
Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.
Date initiated
2024-12-06
Date posted
2025-01-15
Location
Waukesha, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Portrait™ Mobile Monitoring Solution consists of: • Portrait™ Central Viewer Application (Portrait CVAXB) • Portrait™ Core Services (Portrait CSSXB) • Portrait™ Clinical Alarming Unit (Portrait CAU01) • Portrait™ Mobile Patient Monitor Hardware (Portrait HUBXB) • Portrait™ Mobile Patient Monitor Software (Portrait HSWXB) • Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SA01) • Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SP01) • Portrait™ Wearable Pulse O (K234130)Ge Medical Systems Information Technologies, Inc.2024-03-20