Atlas FDA

Polyflux Revaclear Max Dialyzer, Model Number 110634. The Polyflux Revaclear and Polyflux Revaclear Max Dialyzers are si

FDA device recall Z-0490-2015 · posted 2014-12-01 · Terminated

Recall details

Recalling firm
Gambro Renal Products Inc
Reason for recall
Gambro is initiating a field action for Polyflux Revaclear Dialyzer and Polyflux Revaclear Max Dialyzer due to potential tears in the individual packaging that may compromise the sterile barrier.
Action taken
Recall letter was sent via UPS delivery next day to US customers on November 5th, 2014. Product that is returned will be quarantined at the distribution centers.
Root cause
Process control
Status
Terminated
Product code
KDI
Quantity affected
20 million dialyzers
Distribution
Worldwide Distribution: US (nationwide) including PR, and countries of: CA, BB, BM, BS, DO and LC.
Date initiated
2014-11-05
Date posted
2014-12-01
Date terminated
2015-05-20
Location
Opelika, AL