PM1000 Flow Selector - An accessory to an emergency ventilator (resuscitator). The selector is attached to a flowmeter a

FDA device recall Z-1561-2010 · posted 2010-05-09 · Terminated

Recall details

Recalling firm
Precision Medical, Inc.
Reason for recall
The on/off outlet is mislabeled so that when a user/operator turns selector to "on", it actually may be turned "off"
Action taken
The recalling firm issued Customer Alert Product Recall letters on 3/5/10 to inform users of the problem and how to identify affected product.
Root cause
Labeling mix-ups
Status
Terminated
Product code
BTM
Quantity affected
19,518
Distribution
The products were shipped to distributor and medical facilities nationwide. The products were also shipped to Taiwan, Germany, Hong Kong, Canada
Date initiated
2010-03-05
Date posted
2010-05-09
Date terminated
2010-12-13
Location
Northampton, PA