Plastic bottle containing 10.5 mL of IgM Conjugate Product Usage: Product is a used as a counterstain for various indire

FDA device recall Z-1478-2013 · posted 2013-06-05 · Terminated

Recall details

Recalling firm
Bion Enterprises Ltd
Reason for recall
Affected products have a labeled expiration date of one year too long.
Action taken
MBL Bion sent an Urgent Product Recall letter dated January 28, 2013 to all affected customers via email. The letter identified the affected product, problem and actions to be taken. Customers were requested to destroy the affected product and to notify Bion of the number of units consumed and destroyed. For questions call 847-544-5044.
Root cause
Error in labeling
Status
Terminated
Product code
GQH
Quantity affected
201 units
Distribution
Worldwide Distribution - United States (nationwide) in the states of IN, SC and the countries of Canada (Toronto); South Africa (Randburg) and Brazil (Rio de Janeiro)
Date initiated
2013-01-23
Date posted
2013-06-05
Date terminated
2015-05-05
Location
Des Plaines, IL