Atlas FDA

Pioneer Surgical brand Quantum Spinal Rod System Locking Cap; Catalog No 10-LCAP.

FDA device recall Z-1457-05 · posted 2005-08-31 · Terminated

Recall details

Recalling firm
Pioneer Surgical Technology
Reason for recall
The product has a higher than normal locking torque, which may cause the saddle to malfunction or to break interoperatively.
Action taken
Distributors were notified via letter dated July 29, 2005 to retrieve all caps from inventory and point of use locations and to return them to Pioneer.
Root cause
Other
Status
Terminated
Product code
MVH
Quantity affected
5069
Distribution
Nationwide.
Date initiated
2005-07-29
Date posted
2005-08-31
Date terminated
2005-10-25
Location
Marquette, MI

Monitor recalls programmatically