Atlas FDA

Piccolo Chemistry Analyzer software version 6.121 with Electrolyte Panel Lot #8041AC5, model number 400-1022 in polyfoil

FDA device recall Z-2287-2008 · posted 2008-09-17 · Terminated

Recall details

Recalling firm
Abaxis Inc
Reason for recall
The product will recover tCO2 at approximately 5-7 units high relative to some controls and patient samples
Action taken
The firm issued a notification letter to its consignees on April 21, 2008. The firm intends to follow up with phone calls and provide a software upgrade for the products. Contact Abaxis Customer Service at 1-510-675-6500 for assistance.
Root cause
Process control
Status
Terminated
Product code
JLB
Quantity affected
237 kits
Distribution
Product was distributed Worldwide.
Date initiated
2008-04-21
Date posted
2008-09-17
Date terminated
2011-04-26
Location
Union City, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
PICCOLO METLYTE 7 REAGENT DISC (K992140)Abaxis, Inc.1999-08-26