PICC Convenience Kit with Accessories, BioFlo PICC with ENDEXO and PASV Valve Technology, 4F-55 cm SINGLE LUMEN, UPN Pro
FDA device recall Z-0033-2015 · posted 2014-10-07 · Terminated
Recall details
- Recalling firm
- Navilyst Medical, Inc
- Reason for recall
- NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.
- Action taken
- BD Medical sent an Urgent Voluntary Medical Device Recall letter dated July 22, 2014, to all affected customers via Federal Express on August 25, 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to segregate and quarantine all affected product. Immediately discontinue shipment of the affected product. Complete the Recall Response Card form and fax it back to BD at 1-201-847-6990. If the affected product was further distributed cusotmers should identify their cusotmers and notify them immediately of the product recall. Customers with questions were instructed to contact BD Customer Service at 1-888-237-2762, option 3. For questions regarding this recall call 518-742-4571.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- LJS
- Quantity affected
- 60 units/pouches
- Distribution
- Nationwide Distribution
- Date initiated
- 2014-08-25
- Date posted
- 2014-10-07
- Date terminated
- 2017-03-10
- Location
- Glens Falls, NY
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| NMI PICC MAXIMAL BARRIER NURSING KITS (K131038) | Navilyst Medical, Inc. | 2013-05-09 |
| PICC CONVENIENCE KIT (K121990) | Navilyst Medical, Inc. | 2012-08-03 |
| 0.9% SODIUM CHLORIDE INJECTION, USP BD POSTFLUSH SF FLUSH SYRINGE (K042061) | Becton Dickinson | 2004-08-26 |