Atlas FDA

Phoenix x/ray inspector series. Phoenix x|ray inspector series cabinet x-ray systems are used in non-destructive testing

FDA device recall Z-1826-2011 · posted 2011-04-06 · Terminated

Recall details

Recalling firm
GE Inspection Technologies, LP
Reason for recall
Some of the systems do not have a safety interlock on the service access panel. Although the systems have two independent means for indicating when x-rays are being generated, some systems may have one indicator that is not discernible. Also, some systems lack proper placement of a caution label and others do not have this label affixed.
Action taken
The firm has completed all of the following necessary actions and no further actions have been indicated: A Field Service Memo was issued to customers. It described the affected product and described the issue. GE Sensing and Inspection provided and installed retrofit kits, and the systems in question were tested by qualified Field Service Engineers at no charge to customers. GE asked customers to post the notification letter on the affected equipment until retrofit was completed. Questions were to be directed to the GE Service Center at 800-766-6062.
Root cause
Radiation Control for Health and Safety Act
Status
Terminated
Product code
RCE
Quantity affected
118 units total
Distribution
Nationwide Distribution
Date initiated
2009-03-16
Date posted
2011-04-06
Date terminated
2017-04-17
Location
Lewistown, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.