Phoenix x/ray analyser series. Phoenix x|ray analyser series cabinet x-ray systems are used in non-destructive testing,
FDA device recall Z-1825-2011 · posted 2011-04-06 · Terminated
Recall details
- Recalling firm
- GE Inspection Technologies, LP
- Reason for recall
- Some of the systems do not have a safety interlock on the service access panel. Although the systems have two independent means for indicating when x-rays are being generated, some systems may have one indicator that is not discernible. Also, some systems lack proper placement of a caution label and others do not have this label affixed.
- Action taken
- The firm has completed all of the following necessary actions and no further actions have been indicated: A Field Service Memo was issued to customers. It described the affected product and described the issue. GE Sensing and Inspection provided and installed retrofit kits, and the systems in question were tested by qualified Field Service Engineers at no charge to customers. GE asked customers to post the notification letter on the affected equipment until retrofit was completed. Questions were to be directed to the GE Service Center at 800-766-6062.
- Root cause
- Radiation Control for Health and Safety Act
- Status
- Terminated
- Product code
- RCE
- Quantity affected
- 118 units total
- Distribution
- Nationwide Distribution
- Date initiated
- 2009-03-16
- Date posted
- 2011-04-06
- Date terminated
- 2017-04-17
- Location
- Lewistown, PA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.