Philips SureSigns VM Series Patient Monitors: VM4, VM6, VM 8: Model Part Numbers VM4 863063, 453564024371, 453564021961,
FDA device recall Z-2583-2011 · posted 2011-06-17 · Terminated
Recall details
- Recalling firm
- Philips Healthcare Inc.
- Reason for recall
- Speakers on the SureSigns Patient Monitors and Vital Signs, failure may be identified by the display of a Speaker Malfunc message in the alarm message area or an AUDIO FAILED message in the main screen of the device or by the absence of audible sound
- Action taken
- Philips Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated May 2011 to all affected customers. The letter identifies the products, problem, and actions to be taken. The letter states that customers may continue to use the affected products while awaiting replacement of the speaker, provided that the user routinely checks for the display of the "Speaker Malfunct." INOP. Additionally, the letter states that if the customer experience this INOP or there is no sound from the SureSigns Monitor, the customer is to remove the device from use and contact their local Philips service representative. The letter explains that Philips is completing the design for a replacement speaker assembly, which is expected to be available within 8 weeks. When available, Philips will replace the affected speaker assemblies free of charge. Customers are instructed to contact their local Philips Customer Care Center at 1-800-722-9377 if further information or support is required.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- MHX
- Quantity affected
- 515 units
- Distribution
- Worldwide Distribution--USA (nationwide) and countries of ARAB JAMAHIRI, AUSTRIA, AUSTRALIA, BAHRAIN, BANGLADESH, BELGIUM, CANADA ,CHILE, COLOMBIA, CROATIA, CYPRUS, FRANCE, GERMANY, HUNGARY, INDIA, IR
- Date initiated
- 2011-06-10
- Date posted
- 2011-06-17
- Date terminated
- 2013-08-15
- Location
- Andover, MA