Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporatio
FDA device recall Z-1690-2024 · posted 2024-04-25 · Open, Classified
Recall details
- Recalling firm
- Philips North America Llc
- Reason for recall
- Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to the delay in the detection of a change in a patient condition
- Action taken
- Philips issued Urgent Medical Device Correction Letter on 3/22/24. Letter states reason for recall, health risk and action to take: If your UPS fails : 1) Promptly take the UPS offline and plug any devices that had been attached to it into a different appropriate power source. Note: It is recommended that UPS devices intended for operation in network and server rooms be kept in a temperature and humidity controlled environment, ensuring adequate airflow around the UPS to prevent UPS device failure. This notice should be passed on to all those who need to be aware within your organization or to any organization where affected devices have been transferred. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will replace the malfunctioning UPS. If you need any further information, please contact your local Philips representative, or call 1-800- 722- 9377. Philips Reference# 2024-CC-HPM-010
- Root cause
- Component design/selection
- Status
- Open, Classified
- Product code
- MHX
- Quantity affected
- 12,936 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, Estonia, Fiji,
- Date initiated
- 2024-03-22
- Date posted
- 2024-04-25
- Location
- Cambridge, MA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| M3290B Philips IntelliVue Information Center iX (K153702) | Philips Medical Systems | 2016-06-13 |