FDA device recall Z-3287-2011 · posted 2011-09-30 · Terminated

Recall details

Recalling firm
Respironics Novametrix, LLC.
Reason for recall
Pressure lines maybe occluded and display incorrect tidal volumes.
Action taken
Respironics Novametrix, LLC on July 29, 2011 notified accounts via e-mail or tracked receipt, an Urgent - Medical Device Recall letter to notify customers of the issue and to instruct them on conditions for continued use. Monitors will need to be returned for functional performance testing at their facility. Customers can contact Respironics US Customer Service at 1-877-387-3311 for customers residing in the USA or +011 (203)697-6348 for International customers, if they have any questions or need additional information.
Root cause
Manufacturing material removal
Status
Terminated
Product code
BZK
Quantity affected
254 units
Distribution
Worldwide Distribution -- USA, Belgium, Canada, Chile, Columbia, Egypt, India, Italy, Spain, Saudi Arabia Thailand, GB, and Russia.
Date initiated
2011-07-29
Date posted
2011-09-30
Date terminated
2012-05-23
Location
Wallingford, CT