Philips Network Firewall (Cisco ASA 5506), Model # 867098
FDA device recall Z-0850-2018 · posted 2018-03-03 · Terminated
Recall details
- Recalling firm
- Philips Electronics North America Corporation
- Reason for recall
- Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX. The defect involves the clock signal component within the firewall. This component has a high probability of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable. This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot.
- Action taken
- Philips sent an Urgent Recall Notice letter dated October 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that a Philips representatives will contact affected customers to arrange for replacement of all affected units supplied by Philips. For questions contact your local Philips representative: Philips Customer Care Solution Center 800-722-9377.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- MHX
- Quantity affected
- 332 units
- Distribution
- US Nationwide
- Date initiated
- 2017-10-05
- Date posted
- 2018-03-03
- Date terminated
- 2019-07-01
- Location
- Andover, MA