Philips MultiDiagnost Eleva with Digital Image (DI) Rel.2 system. The Philips MultiDiagnost Eleva is intended use as a m

FDA device recall Z-0917-2011 · posted 2011-01-18 · Terminated

Recall details

Recalling firm
Philips Medical Systems North America Co. Phillips
Reason for recall
Images from two different patients may be combined in one patient folder in the Digital Image (DI) or may be combined as one (printed) image. It is possible that an image of a previous patient registered in the DI will be overwritten with an image of a later patient.
Action taken
On March 6, 2009, Philips sent an "URGENT-Device Correction Easy Diagnost Eleva, MultiDiagnost Eleva, and OmniDiagnost Eleva with DI Rel. 2 systems notification" to their consignees/customers. The consignees/customers are advised that before diagnosing images, always check the stored (Digital Image, Viewform or PACS) or printed images to ensure that these represent the data and images captured from the examination of the respective patient. Do not use images if they do not represent the patient that was examined. Consignees will be contacted by Philips to schedule a software upgrade of their units. The consignees can contact Philips Healthcare Call Center at 800-722-9377, #5, #2 and reference "FCO 70800070" with any questions about this recall.
Root cause
Other
Status
Terminated
Product code
IZI
Quantity affected
43 units
Distribution
Nationwide distribution: USA including states of: AZ, CA, CO, CT, DE, FL, GA, HI, ID, LA, MA, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OR, PA, SC, TN, TX, VA, WA, WI, and WY.
Date initiated
2009-03-06
Date posted
2011-01-18
Date terminated
2011-06-16
Location
Bothell, WA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
PHILIPS MULTIDIAGNOST ELEVA (K023441)Philips Medical Systems2002-10-30