Philips MR 7700 System, Model Number 782120

FDA device recall Z-0462-2023 · posted 2022-12-08 · Open, Classified

Recall details

Recalling firm
Philips North America Llc
Reason for recall
Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.
Action taken
An URGENT Medical Device Correction Notice dated 11/22/22 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users 1. Systems that do not yet have the SmokeDetector Interlock feature installed are identified in Appendix A. If your system is listed in Appendix A immediately discontinue use of your impacted MR system(s) until Philips has installed the SmokeDetector Interlock. " Post a Do not use notice on or near the impacted MR system(s) " Post this notice near the affected MR system(s) for ease of reference. " Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved. " Please complete and return the attached customer response form to Philips promptly and no later than 30 days from receipt of this letter. " Note: Once your system has the SmokeDetector Interlock installed, please follow the advice in steps 2 through 5. 2. Philips has installed a SmokeDetector Interlock feature on the systems listed in Appendix B. Immediately confirm if your system has the SmokeDetector Interlock installed by confirming your serial number is listed in Appendix B. " Post this notice near the affected MR system(s) for ease of reference. " Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved. " Please complete and return the attached customer response form to Philips promptly and no later than 30 days from receipt of this letter. 3. The SmokeDetector Interlock feature is integrated with the MR System and powers off the relevant part of the system (Gradient Amplifier) when smoke is detected in order to prevent further development of smoke and/or fire. If the SmokeDetector Interlock has detected smoke and has powered off the Gradient Amplifier: " Remove patient from the system according to the instructions for use because
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
LNH
Quantity affected
566 total
Distribution
US Nationwide.
Date initiated
2022-11-22
Date posted
2022-12-08
Location
Cambridge, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi Nudclei (K213516)Philips Medical Systems Nederland B.V.2022-03-03
Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K193215)Philips Medical Systems Nederland B.V.2020-04-10
Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K183063)Philips Medical Systems Nederland B.V.2019-02-14
Ingenia Elition S and Ingenia Elition X (K173451)Philips Medical Systems Nederland B.V.2018-03-20
Ingenia Elition S and Ingenia Elition X R5.4 (K173356)Philips Medical Systems Nederland B.V.2018-03-06