Philips Medical Systems Computed Tomography X-ray systems (Brilliance CT series, iCT, iCT SP, Ingenuity CT, Ingenuity Co
FDA device recall Z-1722-2014 · posted 2014-06-04 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Reason for recall
- A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.
- Action taken
- Philips Healthcare sent an Urgent Medical Device Correction letter dated April 1, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed If they experience a horizontal, free-floating couch motion, to contact their Field Service Engineer immediately. Customers were instructed to retain a copy of this Field Safety Notice with the equipment Instructions for USER Use (IFU). Customers with questions or needing further information were instructed to contact their local Philips representataive or local Philips Healthcare office. For North America, contact the Customer Care Solutions Center at 1-800-722-9377, Option 5. For questions regarding this recall call 440-483-3032.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- JAK
- Quantity affected
- 6, 839 units
- Distribution
- Worldwide Distribution: US nationwide including PR and DC, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia, Bra
- Date initiated
- 2014-04-01
- Date posted
- 2014-06-04
- Date terminated
- 2018-06-26
- Location
- Cleveland, OH
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| BRILLIANCE VOLUME (K060937) | Philips Medical Systems (Cleveland), Inc. | 2006-06-05 |
| ACQSIM-MULTISLICE-CT, MODEL 453567042271 (K033357) | Philips Medical Systems (Cleveland), Inc. | 2004-01-09 |
| PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM (K033326) | Philips Medical Systems, Inc. | 2003-10-29 |
| MX8000 IDT CT SYSTEM (K012009) | Marconi Medical Systems, Inc. | 2001-09-13 |
| MX8000 V5.0 CT SYSTEM (K010817) | Philips Medical Systems (Cleveland), Inc. | 2001-06-15 |