Philips Medical Systems, Cardiac Viewer or Pulmonary Viewer Application that can run on all Brilliance CT scanners using
FDA device recall Z-0367-2008 · posted 2007-12-29 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Reason for recall
- Incorrect Measurement: Incorrect measurement will occur when the operator uses the distance or area measurement function, such as Graphic Tool line or ROI on zoomed and batch saved images from the Cardiac Viewer or Pulmonary Viewer and then views them in the CT Viewer or on a PACS System.
- Action taken
- Consignees were notified of this recall by an Urgent Device Notification letter sent on 9/10/07. The letter instructed the users to not save any zoomed images to a batch file. Make all measurements of zoomed images in the CT viewer, which provides correct measurements. Do not perform measurements on any zoomed batched images that have been stored on a PACS System. Phillips in investigating the problem and will provide updates free of charge when completed.
- Root cause
- Software design
- Status
- Terminated
- Product code
- JAK
- Quantity affected
- 2,544
- Distribution
- Worldwide Distribution - USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI,
- Date initiated
- 2007-09-10
- Date posted
- 2007-12-29
- Date terminated
- 2012-09-19
- Location
- Cleveland, OH
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.