Philips Medical: 865047 IntelliVue Clinical Information Portfolio Release D.03 865209 IntelliVue Clinical Information Po

FDA device recall Z-3111-2011 · posted 2011-08-31 · Terminated

Recall details

Recalling firm
Philips Healthcare Inc.
Reason for recall
Under certain circumstances, edits to an order can result in unintended scheduled interventions on the IntelliVue Clinical Information Portfolio. A patient may receive interventions or additional medication doses that are not intended.
Action taken
Philips Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated July 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Customers are asked to follow the guidelines in the "Action to be taken by Customer/User" section of the Urgent Medical Device Correction Notice. Philips has developed a software upgrade to address this problem in the short term. This upgrade will be available immediately. This software upgrade will be performed by Philips service personnel. While customers await the correction of their devices, customers are instructed to follow the instructions for use for the device, which includes the directions to: 1. Always use the medication administration record (MAR) in determining patient care. 2. Always review orders and pending administrations in the administration record after any changes to an order are stored. For the longer term, Philips is developing safeguards intended to minimize the potential for similar errors resulting from order changes and upgrade to be available in September 2011. If you need any further information or support concerning this issue, please contact your local Philips representative.
Root cause
Software design
Status
Terminated
Product code
DXJ
Quantity affected
454 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CHINA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY HONG KONG, IRELAND, ITALY, KENYA, LATVIA, MAL
Date initiated
2011-07-08
Date posted
2011-08-31
Date terminated
2020-08-13
Location
Andover, MA