Philips IntelliVue Monitors with software revisions J.21.03, J.21.19. Model Product MP5 M8105A; MP5SC M8105AS; MP5T M810
FDA device recall Z-0104-2015 · posted 2014-10-21 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems, Inc.
- Reason for recall
- Philips Intellivue and Avalon Fetal Monitors in time-synchronized automatic/sequence mode, the NBP automatic measurement series is stopped
- Action taken
- Philips Healthcare issued on 10/3/14 the Urgent Medical Device Correction notification/Field Safety Notice. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. The correction will consist of a software upgrade free of charge. A Philips Healthcare representative will contact customers with affected devices to arrange for service. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: During the interim period until the SW is upgraded please make sure, if the NBP repetition interval needs to be increased (e.g. from 5 to 15 minutes), to restart the automatic NBP measurement. Please review this information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication.
- Root cause
- Software Design Change
- Status
- Terminated
- Product code
- MHX
- Quantity affected
- 7587 units
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of AUSTRALIA AUSTRIA BAHRAIN BELGIUM CHINA COLOMBIA CZECH REPUBLIC DENMARK FINLAND FRANCE GERMANY HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN K
- Date initiated
- 2014-10-03
- Date posted
- 2014-10-21
- Date terminated
- 2017-01-31
- Location
- Andover, MA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| INTELLIVUE CL SPO2 POD AND CL NBP POD, PHILIPS INTELLIVUE PATIENT MONITORS MP5, MP5T, MP5SC (K131913) | Philips Medizin Systeme Boeblingen GmbH | 2013-11-26 |
| INTELLIVUE PATIENT MONITOR (K122439) | Philips Medizin Systeme Boeblingen GmbH | 2012-08-29 |